We recently stocked custom EPA Method 530 standards made specifically for labs participating in Unregulated Contaminant Monitoring Rule 4 (UCMR4). Quantitative Certificates of Analysis with Data Packs are available for each item. Email [email protected] or phone 800-356-1688 ext. 3 to place your order (items are not available at Restek.com).
Cat# 572262 – Method 530 UCMR4 Standard (in Methanol)
Contaminant |
CAS Registry Number |
Minimum Reporting Level |
Standard Concentration |
---|
butylated hydroxyanisole (BHA) |
25013-16-5 |
0.03 µg/L |
150 µg/mL |
o-Toluidine |
95-53-4 |
0.007 µg/L |
35 µg/mL |
Quinoline |
91-22-5 |
0.02 µg/L |
100 µg/mL |
Cat# 572265 – Method 530 UCMR4 Surrogate Standard (in Methanol)
Contaminant |
CAS Registry Number |
Standard Concentration |
---|
Quinoline-D7 |
34071-94-8 |
500 µg/mL |
o-Toluidine-D9 |
194423-47-7 |
500 µg/mL |
Cat# 572266 – Method 530 UCMR4 Internal Standard (in Acetone)
Contaminant |
CAS Registry Number |
Standard Concentration |
---|
Acenaphthene-D10 |
15067-26-2 |
500 µg/mL |
Phenanthrene-D10 |
1517-22-2 |
500 µg/mL |
The suggested analytical column for EPA Method 530, the Determination of Select Semivolatile Organic Chemicals in Drinking Water by Solid Phase Extraction and Gas Chromatography / Mass Spectrometry, is the Rtx-1701, though “any capillary column that provides adequate capacity, resolution, accuracy, and precision may be used.” 3 of the 4 Method 530 target analytes, butylated hydroxyanisole, o-toluidine, and quinoline have been included in UCMR4; dimethipin, the 4th target analyte, was included in UCMR4 as part of EPA 525.3. I’ve included example chromatograms collected at the method reporting limit (MRL) and at 10x the MRL below. The GC inlet, oven and detector settings were taken directly from the method, and the AutoSIM function in MSD Chemstation version D.03 was used to generate the SIM analysis windows.
SIM Analysis of EPA 530 UCMR4 Standard Run at Method Reporting Limit (MRL) on the Rtx-1701
SIM Analysis of EPA 530 UCMR4 Standard Run at 10x the Method Reporting Limit (MRL) on the Rtx-1701
|
Peaks |
tR (min) |
MRL Conc. (µg/mL) |
10x MRL Conc. (µg/mL) |
---|
1. |
o-Toluidine-D9 (SUR) |
8.426 |
1.0 |
1.0 |
2. |
o-Toluidine |
8.481 |
0.070 |
0.70 |
3. |
Quinoline-D7 (SUR) |
10.622 |
1.0 |
1.0 |
4. |
Quinoline |
10.641 |
0.20 |
2.0 |
5. |
Acenaphthene-D10 (IS) |
13.45 |
1.0 |
1.0 |
6. |
Butylated hydroxyanisole (BHA) |
14.25 |
0.30 |
3.0 |
7. |
Phenanthrene-D10 (IS) |
17.318 |
1.0 |
1.0 |
Rtx-1701, 30 m, 0.25 mm ID, 0.25 µm (cat.# 12023) |
Method 530 UCMR4 Standard (cat.# 572262) |
|
Method 530 UCMR4 Surrogate Standard (cat.# 572265) |
|
Method 530 UCMR4 Internal Standard (cat.# 572266) |
|
Dichloromethane |
1 µL pulsed splitless (hold 1.0 min) |
4 mm Single Taper w/Wool (cat.# 23303) |
275 °C |
20 psi (206.8kPa) |
1.05 min |
60 mL/min |
60 °C (hold 1.0 min) to 290 °C at 10 °C/min (hold 1 min) |
He, constant flow |
1.0 mL/min |
MS |
SIM |
Group |
Start Time (min) |
Ion(s) (m/z) |
Dwell (ms) |
---|
1 |
3.50 |
106, 107, 112, 114 |
25 |
2 |
9.55 |
102, 108, 129, 136 |
25 |
3 |
12.08 |
162, 164 |
25 |
4 |
14.02 |
137, 180 |
25 |
4 |
15.92 |
160, 188 |
25 |
|
280 °C |
Quadrupole |
Stainless Steel |
6 mm ID |
280 °C |
180 °C |
3.5 min |
DFTPP |
EI |
HP6890 GC & 5973 MSD |
The EPA Method 530 UCMR4 Standard analyte concentrations vary to simplify preparing ICAL levels based on the minimum method reporting levels. |